NurseCall N46 Wall Transmitter Installation | en 19
Bosch Security Systems User Manual 970.030 | v1.1 | 2009.09
5.7 General configuration
The * marks a factory setting.
(o)
NurseCall N46 Wall Transmitters that feature the output relays are delivered with the Staff
Presence Processing activated (B6 open). Models that do not feature the output relays are
delivered with the Staff Presence Processing disabled (B6 soldered).
Bridge State Function
B3 Open Technical Alarms are not repeated.
Soldered* Technical Alarms are repeated until alarm acknowledgement
(max 20 min.)
B4 Open* Calls (except technical alarms) are repeated until call acknowledge-
ment (max 20 minutes)
Soldered Calls (except technical alarms) are not repeated.
B5 Open* A daily message is sent every 24 hours.
Soldered Daily messages are suppressed.
B6
Open
(o)
Staff Presence Processing is activated.
Soldered
(o)
Staff Presence Processing is disabled.
B7 Open* Activation of J6 [4-5] terminal input causes a Reserve Call.
Soldered Activation of J6 [4-5] terminal input causes a Call for Assistance.
B8 Open* The Assistance Call does not depend on the presence of the Care Per-
sonnel.
Soldered The Assistance Call is possible only if the Care Personnel has signaled
his/her presence.